Manual

LUCAS
TM
2 Chest Compression System – Instructions for Use
100666-00 F Valid from CO J2547, © 2011 JOLIFE AB
30
9.5 Battery physical specifications
9.6 Battery environmental specifications
Category Specifications
Size (H × W × D) 5.1 x 3.5 x 2.2 inches / 13.0 × 8.8 × 5.7 cm
Weight 1.3 lbs / 0.6 kg
Type Rechargeable Lithium Ion Polymer (LiPo)
Capacity 3300 mAh (typical), 86 Wh
Battery voltage (nominal) 25.9 V
Initial Battery runtime (nominal patient) 45 minutes (typical)
Maximum Battery charge time Less than 4 hours at room temperature
(72ºF / 22ºC)
Required interval for replacement of the Battery Recommendation to replace the Battery every 3 years or
after 200 uses (of more than 10 minutes use each time)
Category Specifications
Operating temperature 32ºF to 104ºF / 0ºC to +40ºC, ambient when installed in the
device
Charge temperature 41ºF to 95ºF / 5ºC to +35ºC ambient (68ºF to 77ºF / 20ºC to
25ºC preferred)
Storage temperature 32ºF to 104ºF / 0ºC to +40ºC ambient for less than six
months
IP classification (IEC60529) IP 44