9 Technical specifications
All specifications in this chapter apply to the LUCAS™2 Chest Compression System.
9.1 Patient parameters
9.2 Compression parameters
9.3 Device physical specifications
9.4 Device environmental specifications
Category Specifications
Patients eligible for treatment: Adult patients who fit into the device;
sternum height of 6.7 to 11.9 inches / 170 to 303
mm
a maximum chest width of 17.7 inches / 449 mm
The use of LUCAS is not restricted by patient weight.
Category Specifications
Compression depth (nominal patient) Patients with sternum height over 7.3 inches / 185 mm:
2.1 ±0.1 inches / 53 ±2 mm
Smaller patients with sternum height less than 7.3 inches /
185 mm:
1.5 to 2.1 inches / 40 to 53 mm
Compression frequency 102 ±2 compressions per minute
Compression duty cycle 50 ± 5%
Compression modes (operator selectable) 30:2 (30 compressions followed
by a 3 seconds ventilation pause)
Continuous compressions
Category Specifications
Dimensions when assembled (H × W × D) 22.4 x 20.5 x 9.4 inches / 57 × 52 × 24 cm
Dimensions Carrying Bag with device inside (H × W × D) 25.6 x 13 x 9.8 inches / 65 × 33 × 25 cm
Weight of the device with the Battery 17.2 lbs / 7.8 kg
Category Specifications
Operating temperature +32°F to +104°F / +0°C to +40°C
- 4°F / -20°C for 1 hour after storage at room temperature
Storage temperature -4°F to +158°F / -20°C to +70°C
Relative humidity 5% to 98%, non-condensing
IP classification (IEC60529) IP 43
Operating input voltage 12-24 V DC