2 Chest Compression System – Instructions for Use
100666-00 F Valid from CO J2547, © 2011 JOLIFE AB
28
8.3 Malfunction alarms
Below is a list of all alarms that can occur on LUCAS. You mute all alarms for 60 seconds if you
push MUTE.
Start with manual compressions immediately if LUCAS does not operate properly.
If the malfunction described above seems permanent, LUCAS must be examined by approved
service personnel. Please consult your local Physio-Control representative or JOLIFE AB.
Contact information is available at www.JOLIFE.com
Reason Visual LED indication Audible alarms Result
Compression pattern outside
limit (too deep, too shallow and
timing failure)
Red alarm LED Alarm Compressions stop
Rising temperature in LUCAS Warning alarm None
Too high temperature in LUCAS Red alarm LED Alarm Compressions stop
Hardware error Red alarm LED Alarm Compressions stop
Too high Battery temperature Intermittent red Battery
warning:
The red Battery indication
LED farthest to the right
illuminates intermittently.
Intermittent alarm Compressions stop
Battery charge too low Intermittent red Battery
warning:
The red Battery indication
LED farthest to the right
illuminates intermittently.
Intermittent alarm Compressions stop.
The Battery must be
recharged in the
external Battery
Charger.
Battery error Constant red Battery
warning:
The red Battery indication
LED farthest to the right
shows a constant light.
Alarm Compressions stop.
The Battery cannot be
used anymore.