User Manual

Therapy
4-28 LIFEPAK 20e Defibrillator/Monitor Operating Instructions
5 Monitor screen displays
distortion while pacing.
ECG electrodes not optimally
placed with respect to pacing
electrodes.
Reposition electrodes away
from pacing electrodes.
Patient response to pacing is
highly variable with respect to
capture threshold and ECG
distortion.
Select another lead (I, II, or
III).
Consider changing pacing
rate.
6 Capture does not occur
with pacing stimulus.
Current (mA) set too low. Increase pacing current.
(Administer sedation/
analgesia as needed.)
7
CONNECT ELECTRODES
message appears.
Pacing cable or electrode
disconnected.
Reconnect and set current.
Electrodes not adhering to skin. Prepare skin.
Electrodes outdated. Apply new electrodes and set
current.
8 Pacing stops
spontaneously and
PACER
FAULT
message appears.
Internal error detected. Cycle power and start pacing
again.
Obtain service by qualified
service personnel.
9
REMOVE TEST PLUG
message appears.
Test plug connected to therapy
cable.
Disconnect test plug from
therapy cable.
Connect electrodes to
therapy cable.
10 Intrinsic QRS complexes
not sensed when pacing.
ECG amplitude too low to be
sensed.
Select another lead.
Intrinsic QRS complexes are
occurring during pacemaker's
refractory period.
Adjust pacing rate.
11 Pacing starts
spontaneously.
Patient’s heart rate falls below
set pacing rate.
Appropriate pacemaker
function assess patient.
During standby pacing, ECG
lead disconnects.
Reconnect ECG lead.
12 Set pacing rate (ppm)
and ECG paced rate do
not appear to match.
Internal error detected. Print ECG and calculate the
pace rate.
13 Improper sensing
(for example, sensing on
T-waves).
QRS complex too small.
T-wave too large.
Select another lead.
Table 4-4 Troubleshooting Tips for Noninvasive Pacing (Continued)
Observation Possible Cause Corrective Action