8 Defining Setup Options
Defining Setup Options
LIFEPAK 20e Defibrillator/Monitor Operating Instructions 8-5
©2006-2013 Physio-Control, Inc.
MANUAL MODE SETUP MENU
The Manual Mode Setup menu allows you to define defibrillation and synchronized cardioversion
settings. When you select a menu item, the screen displays a help message. The underlined
options are factory default settings.
Table 8-2 Manual Mode Setup Menu
Menu Item Help Message Options
SYNC
Set up sync defaults Refer to Table 8-3.
PADDLES
DEFAULT
Default energy for paddles
or QUIK-COMBO
electrodes
Power-on energy setting for standard
paddles and therapy electrodes:
2, 5, 10,
50, 100, 125, 150, 175, 200
(joules), or
ENERGY PROTOCOL.
ENERGY
PROTOCOL...
Energies for energy
protocol
Refer to Table 8-4.
INTERNAL
DEFAULT
Default energy for internal
paddles
Power-on energy setting for internal
paddles:
2, 5, 10, 20, 30, or 50.
VOICE PROMPTS
Voice prompts active in
manual mode
ON: Voice prompts active.
OFF: Voice prompts inactive.
SHOCK TONE
Tone when defibrillator is
fully charged
ON: A tone sounds.
OFF: No tone.
MANUAL ACCESS
Means for accessing
manual mode
MANUAL / DIRECT: No restrictions to
manual mode.
AED / DIRECT: No restrictions to AED
mode.
AED / CONFIRM: Confirmation required to
gain manual access.
AED / PASSCODE: Passcode required to
enter manual mode.
SET PASSCODE...
Passcode required to
enter manual mode
If configured for passcode access:
None: Default passcode enabled.
New: User-defined 4-digit code enabled.
Table 8-3 Synchronization Defaults
Menu Item Help Message Options
SYNC AFTER
SHOCK
Resume sync after energy
transfer
ON: Defibrillator returns to synchronous
mode after energy transfer.
OFF: Defibrillator returns to asynchronous
mode.
REMOTE SYNC
Allow sync with remote
monitor
ON: Remote synchronization active with
connection to a compatible remote monitor.
OFF: No remote monitor connection active.