User Manual

Maintaining the Equipment
7-6 LIFEPAK 20e Defibrillator/Monitor Operating Instructions
Patient ECG Cable Check
Equipment needed:
LIFEPAK 20e defibrillator/monitor
Fully charged batteries
Patient ECG cable
(3-wire or 5-wire)
3-lead or 12-lead simulator
Procedure:
1 Press
ON.
2 Connect the ECG cable to the defibrillator.
3 Connect all cable leads to the simulator.
4 Turn on the simulator and select a rhythm.
5 After a few seconds confirm that the screen displays a rhythm and no
LEADS OFF or SERVICE
messages appear.
Standard Paddles Monitoring Check
Equipment needed:
LIFEPAK 20e defibrillator/monitor
Standard paddles
Procedure:
1 Press
ON.
2 Select
PADDLES lead.
3 Press the paddle electrode surfaces together and confirm that a flat line appears.
4 Shake each paddle in the air and confirm that irregular noise signals appear.
5 Install the paddles in the paddle wells.
Standard Paddles Defibrillation and Synchronized Cardioversion Check on
Battery Power
Equipment needed:
LIFEPAK 20e defibrillator/monitor with fully charged internal battery
Standard paddles
Defibrillator checker
Patient ECG cable
3-lead or 12-lead patient simulator
WARNING!
Shock hazard.
When discharged as described in this test, the defibrillator delivers up to 360 J of electrical
energy. Unless discharged properly as described in this test, this electrical energy may cause
serious personal injury or death. Do not attempt to perform this test unless you are qualified by