Owner manual

Therapy
4-24 LIFEPAK 20e Defibrillator/Monitor Operating Instructions
8 ABNORMAL ENERGY
DELIVERY
message appears
and Shock XJ Abnormal
annotated on printout.
Open air discharge with
standard paddles.
Press paddles firmly on
patient’s chest when
discharging.
Perform test discharges with
defibrillation checker.
Discharge occurs with
standard paddles shorted
together.
Refer to warning, page 4-15.
Patient impedance out of
range.
Increase energy and/or
repeat discharges as
needed.
Consider replacing
disposable therapy
electrodes with new.
9
CONNECT ELECTRODES
message appears.
Inadequate connection to
defibrillator.
Check for electrode
connection.
Electrodes do not adhere
properly to the patient.
Press electrodes firmly on
patient’s skin.
Clean, shave, and dry the
patient’s skin as
recommended.
Electrodes are dry, damaged,
or out of date.
Replace the electrodes.
10
CONNECT TO AC POWER
message appears.
Remote sync is selected and
the defibrillator is not
connected to AC power.
Connect to AC power.
Press
SYNC to turn off remote
sync.
11
CONNECT SYNC CABLE TO
REMOTE MONITOR
message
appears.
Remote sync is selected and
the defibrillator is not
connected to the remote
monitor.
Connect to remote monitor.
Press
SYNC to turn off remote
synchronization.
12
LOW IMPEDANCE–
RECHARGING
message
appears.
Patient impedance of <15
ohms detected.
No action required.
13
SEARCHING FOR SIGNAL
message appears.
Remote sync is selected and
the defibrillator is qualifying
the input signal.
No action required.
Table 4-3 Troubleshooting Tips for Defibrillation and Synchronized Cardioversion (Continued)
Observation Possible Cause Corrective Action