Owner manual

8 Setup Options
Setup Options
LIFEPAK 20e Defibrillator/Monitor Operating Instructions 8-7
©2006-2013 Physio-Control, Inc.
AED MODE SETUP MENU
The AED Mode Setup menu allows you to define automated external defibrillator (AED) settings.
When you select a menu item, the screen displays a help message describing the option. The
underlined options are factory default settings and are consistent with 2010 American Heart
Association (AHA) and European Resuscitation Council (ERC) guidelines. Refer to Appendix F
for a more detailed description of CPR setup options.
Table 8-5 AED Mode Setup Menu
Menu Item Help Message Options
ENERGY
PROTOCOL...
Sequence of defibrillation
energies
Refer to Table 8-7.
VOICE
PROMPTS
Voice prompts on in AED
Mode
ON: Voice prompt active.
OFF: Voice prompt inactive.
AUTO ANALYZE
Select Auto Analyze options AFTER 1ST SHOCK: The second and third
rhythm analyses of each three-shock stack
start automatically. (Stacked Shocks must
be set to
ON.)
OFF: Auto analyze inactive.
MOTION
DETECTION
Alert when motion is detected ON or OFF.
ECG DISPLAY
Display ECG waveform in
AED Mode.
ON or OFF.
CPR...
Set CPR options for AED
Mode
PULSE CHECK
Enable Pulse Check prompt ALWAYS: After every stack of shocks and
every NSA finding
.
AFTER EVERY NSA
: Only after NO SHOCK
ADVISED.
AFTER SECOND NSA
: After every NSA
except for first analysis NSA result.
NEVER: Never prompt for PULSE CHECK.
Table 8-6 AED Mode CPR Setup Mode
Menu Item Help Message Options
CPR TIME 1
Set CPR interval after shocks 15, 30, 45, 60, 90, 120, 180 SECONDS, or
30 MINUTES
CPR TIME 2
Set CPR interval after NO
SHOCK ADVISED
15, 30, 45, 60, 90, 120, 180 SECONDS, or
30 MINUTES
INITIAL CPR
Enable Initial CPR OFF, ANALYZE FIRST, CPR FIRST
INITIAL CPR
TIME
Set CPR interval for initial
CPR