Owner manual

Setup Options
8-6 LIFEPAK 20e Defibrillator/Monitor Operating Instructions
To activate, select ENERGY PROTOCOL in the Paddle Default menu. Auto energy sequences
are disabled if you press the
ENERGY SELECT control or change to or from AED mode during
use.
Table 8-4 Manual Mode Energy Protocol Setup Menu
Menu Item Help Message Options
PRESET
PROTOCOL
Select preset energy
protocol
FULL RANGE, PEDIATRIC.
ENERGY 1
Select energy level for
shock 1
Full range: 100, 125, 150, 175, 200, 225,
250, 275, 300, 325, 360.
Pediatric: 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30,
50, 70, 100
.
ENERGY 2*
Select energy level for
shock 2
Full range: 100, 125, 150, 175, 200, 225,
250, 275
, 300, 325, 360.
Pediatric: 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30,
50, 70, 100.
ENERGY 3*
Select energy level for
shock 3
Full range: 100, 125, 150, 175, 200, 225,
250, 275, 300, 325
, 360.
Pediatric:
2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30,
50, 70, 100, 125, 150.
*ENERGY 2 cannot be less than ENERGY 1. ENERGY 3 cannot be less than ENERGY 2.