7 Maintaining the Equipment
Maintaining the Equipment
LIFEPAK 20e Defibrillator/Monitor Operating Instructions 7-11
©2006-2013 Physio-Control, Inc.
11 Date printed on report
is incorrect.
Date is incorrectly set. Change the date setting. Refer to
Section 2, page 2-7.
12 Displayed messages
are faint or flicker.
Low battery power.
Out of temperature range.
Connect to AC power immediately.
13 Low speaker volume. Moisture in speaker grill
holes.
Wipe moisture from speaker grill and
allow device to dry.
14
MAINTENANCE DUE
message appears.
Maintenance prompt is set
to display at a selected
interval in service mode.
Continue to use device if needed.
Contact service personnel to reset or
turn off the maintenance prompt.
Contact Physio-Control Technical
Support for instructions on how to
reset or turn off this prompt.
15
SELF TEST DID NOT
COMPLETE
message
appears.
Test plug not connected to
QUIK-COMBO therapy
cable during daily auto test.
Connect test plug to QUIK-COMBO
therapy cable.
Perform manual user test.
Standard paddles not
seated in paddle wells
during daily auto test.
Make sure standard paddles are
securely seated in the paddle wells
for daily auto test.
Defective therapy cable or
a problem with the
defibrillator.
Contact qualified service personnel.
16
SELF TEST FAILED
message appears.
Device self-test circuitry
detects service condition
during daily auto test.
Use defibrillator or pacemaker if
needed in an emergency.
Report occurrence of self-test failure
to qualified service personnel.
17
USER TEST DID NOT
COMPLETE
message
appears.
Test plug not connected to
QUIK-COMBO therapy
cable during user test.
Connect test plug to QUIK-COMBO
therapy cable for user test.
Standard paddles not
seated in paddle wells
during user test.
Make sure standard paddles are
securely seated in the paddle wells
for user test.
Defective therapy cable or
a problem with the
defibrillator.
Contact qualified service personnel.
18
USER TEST FAILED
message appears.
Device self-test circuitry
detects service condition
during user test.
Use defibrillator or pacemaker if
needed in an emergency.
Report occurrence of user-test
failure to qualified service personnel.
19 Printout annotated
79 J, but > 79 J was
selected.
Paddles discharged when
seated in the paddle wells
or shorted together.
Refer to "Standard Paddles
Defibrillation and Synchronized
Cardioversion Check on Battery
Power," page 7-6.
Perform User Test.
20 Defibrillator does not
charge to 200 J within
10 seconds while on
battery power.
Low battery condition.
Defective battery.
Connect the defibrillator to AC
power.
Contact qualified service personnel.
Table 7-2 General Troubleshooting Tips (Continued)
Observation Possible Cause Corrective Action