4-44
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Monitoring Noninvasive Blood Pressure
For general troubleshooting tips, see Table 10-2 on page 10-18.
NIBP TIME OUT message appears The monitor did not complete a
measurement in 120 seconds
Check cuff for snug fit on
patient.
Check that cuff artery
markings are aligned with
the artery.
Repeat measurement.
Try a higher initial pressure.
If this message persists, use
another method to measure
the patient’s blood pressure.
NIBP WEAK PULSE message
appears
The monitor did not detect any
pulses
Check pulses distal to the
cuff.
Check cuff for snug fit on
patient.
Check that cuff artery
markings are aligned with
the artery.
XXX appears in place of
NIBP readings
NIBP module failed.
NIBP module failed to calibrate
successfully.
Turn device off and then on
again.
If problem persists, contact
qualified service personnel.
NIBP CHECK CUFF message
appears
Cuff is not connected to patient
or device
Check cuff for snug fit on
patient.
Check that cuff artery
markings are aligned with
the artery.
Check cuff tubing
connection to device.
Unable to connect NIBP tubing
to device
The LIFEPAK 12 NIBP tubing
connector is not compatible
with the LIFEPAK 15 NIBP port
Obtain correct NIBP tubing
that is compatible with
LIFEPAK 15 monitor/
defibrillator.
Cuff not deflating Internal valves fail to open Disconnect NIBP tubing.
Remove cuff from patient.
Cuff not inflating Cuff is not connected to the
device
Check tubing connection to
device and cuff.
Leak in tubing, cuff, or
connector
Replace NIBP tubing or
cuff.
Table 4-5 Troubleshooting Tips for NIBP Monitoring (Continued)
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION