Troubleshooting Tips
Table 4-5 Troubleshooting Tips for NIBP Monitoring
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION
NIBP AIR LEAK message appears Cuff applied too loosely
Leak in cuff/monitor pneumatic
system
Check cuff for snug fit on
patient.
Check that the cuff/monitor
connection is secure.
Check cuff for leaks. Do not
use a cuff that exhibits a
leak.
NIBP FLOW ERROR message
appears
The pneumatic system is not
maintaining stable cuff
pressure
Deflate or remove cuff.
Check tubing for leaks.
Replace cuff.
NIBP FAILED message appears The monitor cannot establish
zero-pressure reference
Check tubing for kink or
blockage.
If this message persists,
remove monitor from use
and obtain service. Use
another method to measure
the patient’s blood pressure.
NIBP INITIALIZING message
appears
NIBP requested while NIBP
module is still initializing
Wait until message
disappears and request
NIBP.
NIBP MOTION message appears The patient extremity moved
too much for the monitor to
accurately complete the
measurement
Have patient lie quietly with
extremity relaxed and
supported.
Check that patient’s arm
does not move during NIBP
measurement.
NIBP OVERPRESSURE message
appears
Cuff pressure exceeded
290 mmHg
Disconnect tubing or remove
cuff.
Avoid very rapid squeezing
of the cuff.
If this message persists,
remove the cuff from use
and obtain service.