Manual

MONITORING
4
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
4-35
SPO2: CHECK SENSOR message
appears
Sensor is disconnected from
patient or cable
Attach the sensor.
Check that sensor is secure.
Excessive ambient light Remove or block light
source, if possible.
Cover sensor with opaque
material, if necessary.
Faulty or defective sensor Replace sensor.
Patient has a weak pulse or low
blood pressure, or the sensor is
not properly placed
Change sensor location.
Check if patient perfusion is
adequate for sensor
location.
Check that sensor is secure
and not too tight.
Check that sensor is not on
extremity with blood
pressure cuff or
intravascular line.
Test sensor on someone
else.
SPO2: UNKNOWN SENSOR
message appears
A sensor that is not Physio-
Control approved is connected
to the device
Check that the sensor is
approved by Physio-Control.
If using Nellcor sensor,
check that it is connected to
monitor using
Masimo
Red MNC adapter cable.
SPO2: SEARCHING FOR PULSE
message appears
A sensor is connected to the
patient and is searching for a
pulse
Wait for completion.
SPO2: LOW PERFUSION message
appears
Patient has a weak pulse Change sensor location.
SP02: POOR QUALITY SIGNAL
message appears
When the signal quality is low,
the accuracy of the
measurement may be
compromised
Check that sensor and cable
are connected properly.
Move sensor to a better
perfused site.
SPCO: POOR QUALITY SIGNAL
message appears
When the signal quality is low,
the accuracy of the
measurement may be
compromised
Check that sensor and cable
are connected properly.
Move sensor to a better
perfused site.
SPMET: POOR QUALITY SIGNAL
message appears
When the signal quality is low,
the accuracy of the
measurement may be
compromised
Check that sensor and cable
are connected properly.
Move sensor to a better
perfused site.
Table 4-4 Troubleshooting Tips for SpO
2
, SpCO, and SpMet (Continued)
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION