Manual

4-14
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Monitoring the ECG
For general troubleshooting tips, see Table 10-2 on page 10-18.
ECG amplitude too small Poor electrode-skin contact Prepare skin and apply new
electrodes.
ECG lead selected Increase ECG gain or change ECG
lead.
Patient condition (for example,
significant myocardial muscle
loss or tamponade)
Increase ECG gain or change ECG
lead.
Monitor displays dashed
lines with no ECG
LEADS
OFF
messages
PADDLES lead selected but
patient connected to ECG
cable
Select one of the limb or
precordial leads.
Monitor shows isoelectric
(flat) line and
PADDLES lead
selected
The Test Load is connected to
therapy cable
Remove the Test Load and
connect therapy electrodes to
cable.
Connect ECG cable and select
another lead.
Internal pacemaker pulses
difficult to see
Pacemaker pulses are very
small
Turn on internal pacemaker
detector (see "Monitoring Patients
Who Have Internal Pacemakers"
on page 4-11).
Monitor frequency response
limits visibility
Connect ECG cable and select a
lead other than
PADDLES.
Print ECG in Diagnostic mode
(see "How to Print a Current
Report" on page 7-10).
Table 4-2 Troubleshooting Tips for ECG Monitoring (Continued)
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION