Manual

MONITORING
4
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
4-13
Monitor displays dashes
(---) instead of heart rate
Heart rate is < 20 bpm Use ECG printout to calculate
heart rate.
Heart rate is > 300 bpm Use ECG printout to calculate
heart rate.
Pacing function is active No corrective action needed.
Poor ECG signal quality Poor electrode-skin contact Reposition cable or lead wires to
prevent electrodes from pulling
away from patient.
Secure trunk cable clasp to
patient’s clothing.
Prepare skin and apply new
electrodes.
Outdated, corroded, or dried-
out electrodes
Check Use By date on electrode
packages.
Use only unexpired silver/silver
chloride electrodes. Leave
electrodes in sealed pouch until
time of use.
Loose connection.
Damaged cable or connector/
lead wire
Check or reconnect cable
connections.
Inspect ECG and therapy cables.
Replace if damaged.
Check cable with simulator and
replace if malfunction observed.
Noise because of radio
frequency interference (RFI)
Check for equipment causing RFI
(such as a radio transmitter) and
relocate or turn off equipment
power.
Baseline wander
(low frequency/high
amplitude artifact)
Inadequate skin preparation Prepare skin and apply new
electrodes.
Poor electrode-skin contact Check electrodes for proper
adhesion.
Diagnostic frequency response Print ECG in monitor frequency
response.
Fine baseline artifact (high
frequency/low amplitude)
Inadequate skin preparation Prepare skin and apply new
electrodes.
Isometric muscle tension in
arms/legs
Confirm that limbs are resting on
a supportive surface.
Check electrodes for proper
adhesion.
Table 4-2 Troubleshooting Tips for ECG Monitoring (Continued)
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION