Manual

4-12
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Monitoring the ECG
Troubleshooting Tips
If problems occur while monitoring the ECG, check Table 4-2 for aid in troubleshooting. For basic
troubleshooting problems, such as no power, see "General Troubleshooting Tips" on page 10-18.
Table 4-2 Troubleshooting Tips for ECG Monitoring
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION
Any of these messages
displayed:
CONNECT ELECTRODES
CONNECT ECG LEADS
ECG LEADS OFF
XX LEADS OFF
Therapy electrodes not
connected
Connect therapy electrode.
One or more ECG electrodes
disconnected
Connect ECG electrode.
ECG cable is not connected to
monitor
Connect ECG cable.
Poor electrode-skin contact Reposition cable or lead wires to
prevent electrodes from pulling
away from patient.
Secure trunk cable clasp to
patient’s clothing.
Prepare skin and apply new
electrodes.
PACER was pressed. The
monitor automatically
switched to Lead II, but ECG
leads are not connected.
Connect ECG leads and initiate
pacing.
Broken ECG cable lead wire Select another lead.
Select
PADDLES lead, and use
standard paddles or therapy
electrodes for ECG monitoring.
Check ECG cable continuity.
Screen blank and
ON LED
illuminated
Screen not functioning
properly
Print ECG on recorder as backup.
Contact service personnel for
repair.
Systole beeps not heard or
do not occur with each
QRS complex
Volume too low Adjust volume.
QRS amplitude too small to
detect
Adjust ECG size.
Displayed heart rate (HR)
different than pulse rate
ECG size set too high or too
low
Adjust ECG size
up or down.
Monitor detecting the patient’s
internal pacemaker pulses
Change monitor lead to reduce
internal pacemaker pulse size.
Displayed heart rate (HR)
different from displayed
ECG waveform
ECG size set too high or too
low
Adjust ECG size
up or down.
Monitor detecting the patient’s
internal pacemaker pulses
Change monitor lead to reduce
internal pacemaker pulse size.