LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Screen Messages
SPO2: SEARCHING FOR PULSE A sensor is connected to the patient and is searching for a
pulse.
SPO2: SENSOR DOES NOT
SUPPORT SPCO OR SPMET
The sensor in use only measures SpO
2
.
SPO2: UNKNOWN SENSOR A sensor that is not Physio-Control approved is connected to the
device.
STAND CLEAR/PUSH SHOCK
BUTTON
Prompts you to stand clear and push (shock).
START CPR Prompts you to begin providing CPR to the patient.
SWITCHING PRIMARY TO LEAD II Pacing is turned on while
PADDLES is the primary lead.
SWITCHING PRIMARY TO
PADDLES
Device was in Lead II when
ANALYZE was pressed. PADDLES
becomes the primary lead.
SYNC MODE Device is currently in Sync mode.
TEMP: ACCURACY OUTSIDE
LIMITS
Temperature accuracy check has failed.
TEMP: CHECK SENSOR Device is not receiving sufficient input from sensor.
TO CANCEL, PUSH SPEED DIAL The defibrillator is charging or charged and the device may be
disarmed by pressing the Speed Dial.
TRANSMISSION CANCELLED Data transmission has been cancelled.
TRANSMISSION COMPLETED Data transmission completed successfully.
TRANSMISSION FAILED Data transmission was not successful.
TRANSMITTING TO <SITE> Connection is established to <site> and transmission of
requested report is occurring.
UNABLE TO CONNECT Unable to establish connection with Bluetooth device.
UNABLE TO TRANSMIT Unable to send data.
UNKNOWN DEVICE Bluetooth connection failed or timed out before obtaining target
device name.
USE ECG LEADS Sync mode attempted, but ECG electrodes are not attached to
patient,
PADDLES lead is displayed, and standard paddles are
connected to defibrillator.
USER TEST FAILED Unsuccessful User Test.
USER TEST IN PROGRESS
USER TEST selected on the OPTIONS menu and test is in process.
USER TEST PASSED Successful User Test completed.
VX LEADS OFF ECG electrode such as “V1” is disconnected.
X DEVICES FOUND Shows number of Bluetooth-enabled devices found.
Table B-1 Summary of Screen Messages (Continued)
MESSAGE DESCRIPTION