A-16
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Specifications and Performance Characteristics
Visual Alarms Alarms are indicated visually by:
• The violated parameter flashes in inverse video with a message in the message
area of the display.
• These visual indications remain on the display until the alarm is corrected. Visual
indication of alarms continue even when the tones have been silenced.
Table A-5 Setup Options Factory Default Settings
MENU MENU/ITEM FACTORY DEFAULT SETTINGS
General Language (Country Specific)
Code Summary Long
Trend Summary Off
Site Number 000
Device ID
“LP15” + last 4 digits of serial number, for example,
LP151234
Auto Log On
Line Filter 60 Hz
Timeout Speed 30 seconds
Manual mode Sync After Shock Off
Pads Default 200 (joules)
Energy Protocol Inactive
Internal Default 10 (joules)
Voice Prompts On
Shock Tone On
Manual Access Manual / Direct
Set Passcode 0000
AED mode Energy Protocol 200–300–360
Auto Analyze Off
Motion Detection On
Pulse Check Never
CPR CPR Time 1 120 seconds
CPR Time 2 120 seconds
Initial CPR Off
Initial CPR Time 120 seconds
Preshock CPR Off
CPR Metronome Metronome On
Adult - No Airway 30:2
Adult - Airway 10:1
Youth - No Airway 15:2
Youth - Airway 10:1
Table A-4 Alarm Performance Characteristics (Continued)
CHARACTERISTIC DESCRIPTION