LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Specifications and Performance Characteristics
Pre-Shock CPR Allows the user to be prompted for CPR while the device is charging.
Options are
OFF, 15, 30 seconds
Pulse Check Allows the user to be prompted for a pulse check at various times.
Options are
ALWAYS, AFTER SECOND NSA, AFTER EVERY NSA, NEVER
Stacked Shocks Allows for CPR after 3 consecutive shocks or after a single shock. Options
are
OFF, ON
CPR Time 1 or 2 User selectable times for CPR. Options are 15, 30, 45, 60, 90, 120, 180
seconds and
30 minutes
PACER
Pacing Mode Demand or nondemand
Rate and current defaults
Pacing Rate 40 to 170 PPM
Rate Accuracy ±1.5% over entire range
Output Waveform Monophasic, truncated exponential current pulse (20 ±1.5 msec)
Output Current 0 to 200 mA
Pause: Pacing pulse frequency reduced by a factor of 4 when activated
Refractory Period 180 to 280 msec (function of rate)
ENVIRONMENTAL—Unit meets functional requirements during exposure to the following environments
unless otherwise stated.
Operating Temperature 0° to 45°C (32° to 113°F)
-20°C (-4°F) for 1 hour after storage at room temperature
60°C (140°F) for 1 hour after storage at room temperature
Storage Temperature -20° to 65°C (-4° to 149°F) except therapy electrodes and batteries
Relative Humidity,
Operating
5 to 95%, non-condensing
NIBP: 15 to 95%, non-condensing
Relative Humidity,
Storage
10 to 95%, non-condensing
Atmospheric Pressure,
Operating
-382 to 4,572 m (-1,253 to 15,000 ft)
NIBP: -152 to 3,048 m (-500 to 10,000 ft)
Water Resistance,
Operating
IP44 (dust and splash resistance) per IEC 529 and EN 1789 (without
accessories except for 12-lead ECG cable, hard paddles, and battery
pack)
Vibration MIL-STD-810E Method 514.4
Propeller Aircraft - category 4 (figure 514.4-7 spectrum a)
Helicopter - category 6 (3.75 Grms)
Ground Mobile - category 8 (3.14 Grms)
EN 1789: Sinusoidal Sweep, 1 octave/min, 10-150 Hz, ±0.15 mm/2 g
Shock (drop) 5 drops on each side from 18 inches onto a steel surface
EN 1789: 30-inch drop onto each of 6 surfaces
Shock (functional) Meets IEC 60068-2-27 and MIL-STD-810E shock requirements
3 shocks per face at 40 g, 6 ms half-sine pulses