LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Specifications and Performance Characteristics
DEFIBRILLATOR
Charge Time (per EN 60601-2-4)
AC Operation Only:
Maximum Time from Charge to Shock Ready (Manual Mode):
Voltage Charge Time
90-240 Vac (Nominal): 360 J within 10 seconds
Maximum Time from Initiation of Analysis to Shock Ready (AED Mode):
Voltage Charge Time
90-240 Vac (Nominal) 360 J within 30 seconds
Maximum Time from Power-on to Shock Ready (Manual Mode):
Voltage Charge Time
90-240 Vac (Nominal) 360 J within 25 seconds
Maximum Time from Power-on to Shock Ready (AED Mode):
Voltage Charge Time
90-240 Vac (Nominal) 360 J within 40 seconds
DC Operation Only:
Maximum Time from Charge to Shock Ready (Manual Mode):
Voltage Charge Time
11-17.6 Vdc (Nominal) 360 J within 10 seconds
Maximum Time from Initiation of Analysis to Shock Ready (AED Mode):
Voltage Charge Time
11-17.6 Vdc (Nominal) 360 J within 30 seconds
Maximum Time from Power-on to Shock Ready (Manual Mode):
Voltage Charge Time
11-17.6 Vdc (Nominal) 360 J within 25 seconds
Maximum Time from Power-on to Shock Ready (AED Mode):
Voltage Charge Time
11-17.6 Vdc (Nominal) 360 J within 40 seconds
Battery Operation Only:
Maximum Time from Charge to Shock Ready (Manual Mode):
Battery Status Charge Time
Fully charged 200 J within 7 seconds, nominal
Fully charged, followed by 15
full-energy shocks
360 J within 10 seconds
Fully charged 360 J within 10 seconds
Table A-1 LIFEPAK 15 Monitor/Defibrillator Specifications (Continued)