LIFEPAK 15 Monitor/Defibrillator Operating Instructions
A-7
TEMPERATURE
Sensors Measurement Specialties 4400 series esophageal/rectal and Foley
catheter temperature probes, and 4499HD skin temperature probe
Displayed Range 24.8° to 45.2°C (76.6° to 113.4°F)
Resolution 0.1°C
Accuracy ±0.2°C
Labels Temp, T-esoph, T-naso, T-bladder, T-rectal, T-skin
Update Rate Every 10 seconds, minimum
Adapter Cable Only use Physio-Control part number 3303935
Cable Length 1.5 or 3 m (5 or 10 ft)
TREND
Time Scale Auto, 30 minutes, 1, 2, 4, or 8 hours
Duration Up to 8 hours
ST After initial 12-lead ECG analysis, automatically selects and trends ECG
lead with the greatest ST displacement
Display Choice of HR, PR (SpO
2)
, PR (NIBP), SpO
2
(%), SpCO(%), SpMet(%),
CO
2
(EtCO
2
/FiCO
2)
, RR (CO
2
), NIBP, IP1, IP2, TEMP, ST
ALARMS
Quick Set Activates alarms for all active vital signs
VF/VT Alarm Activates continuous CPSS monitoring in Manual mode
Apnea Alarm Occurs when 30 seconds has elapsed since last detected respiration
Heart Rate Alarm Limit
Range
Upper, 100–250 bpm; lower, 30–150 bpm
INTERPRETIVE
ALGORITHM
12-Lead Interpretive Algorithm: University of Glasgow 12-Lead ECG
Analysis Program, includes AMI and STEMI statements
PRINTER
Prints continuous strip of the displayed patient information and reports
Paper Size 100 mm (3.9 in)
Print Speed 25 mm/sec or 12.5 mm/sec
Optional 50 mm/sec time base for 12-lead ECG reports
Delay 8 seconds
Autoprint Waveform events print automatically
Frequency Response Diagnostic—0.05 to 150 Hz or 0.05 to 40 Hz
Monitor—0.67 to 40 Hz or 1 to 30 Hz
Table A-1 LIFEPAK 15 Monitor/Defibrillator Specifications (Continued)
CHARACTERISTIC DESCRIPTION