MAINTAINING THE EQUIPMENT
10
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
10-21
CHECK PRINTER message
appears
Printer paper jams, slips, or
misfeeds
Reinstall paper.
If problem persists, contact
qualified service technician.
Printer is out of paper Add new paper.
Service LED illuminates Device self-test circuitry
detects service condition
Continue to use defibrillator or
pacemaker, if needed.
Turn device off and then on again.
Note that this creates a new
“patient.” If Service LED does not
clear, remove device from active
use.
Report occurrence of Service LED to
qualified service personnel.
Obtain another defibrillator, if
necessary.
ECG monitoring problems See "Troubleshooting Tips" on
page 4-12.
Problems with AED
operation
See "Troubleshooting Tips" on
page 5-19.
Problems with
defibrillation/synchronized
cardioversion
See "Troubleshooting Tips" on
page 5-29.
Problems with pacing See "Troubleshooting Tips" on
page 5-36.
Displayed time is incorrect Time is incorrectly set Change the time setting. See
"Options" on page 3-23.
Date printed on report is
incorrect
Date is incorrectly set Change the date setting. See
"Options" on page 3-23.
Displayed messages are
faint or flicker
Low battery power
Out of temperature range
Replace the battery immediately.
Connect to auxiliary power using
approved power adapter.
Low speaker volume Moisture in speaker grill
holes
Wipe moisture from speaker grill and
allow device to dry.
MAINTENANCE DUE message
appears
Maintenance prompt is set
to display at a selected
interval in Service mode
Continue to use device, if needed.
Contact service personnel to perform
routine maintenance.
Contact Physio-Control Technical
Support for instructions on how to
reset or turn off this prompt.
Table 10-2 General Troubleshooting Tips (Continued)
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION