LIFEPAK 15 Monitor/Defibrillator Operating Instructions
General Troubleshooting Tips
Fuel gauge on battery does
not illuminate
Extremely depleted battery Charge battery in Station-Mobile or
REDI-CHARGE battery charger.
Faulty battery Replace battery.
Device turns off
unexpectedly
High power draw Press
ON immediately to turn device
back on.
Low battery power Replace battery immediately.
Press
ON to turn device back on.
RF equipment too close to
defibrillator
Separate RF equipment from
defibrillator. See "Separation
Distances" on page D-4.
Press
ON to turn device back on.
Cellular equipment too close
to installed battery
Move cellular equipment away from
installed battery.
Press
ON to turn device back on.
If device does not turn on, replace
battery.
LIFENET Gateway (modem)
too close to installed battery
Store modem in side pouch of
defibrillator. Do not store modem in
back pouch.
Press
ON to turn device back on.
If device does not turn on, replace
battery.
Device won’t turn off
ON not pressed long enough
to turn off device
Press and hold ON for at least two
seconds.
Monitor/defibrillator
operates, but screen is
blank
Operating temperature is too
low or too high
Operate defibrillator within specified
ambient temperature range.
Screen not operating
properly
Print ECG strip to assess rhythm and
other active vital signs.
Press
ANALYZE and use AED mode, if
necessary.
Contact qualified service technician.
Monitor/defibrillator
operates, but screen not
readable
Screen in direct sunlight Change screen from color to black
and white.
Reposition or shield device.
Print ECG strip to assess rhythm and
other active vital signs.
Press
ANALYZE and use AED mode, if
necessary.
Table 10-2 General Troubleshooting Tips (Continued)
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION