MAINTAINING THE EQUIPMENT
10
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
10-19
AUXILIARY POWER LED not
illuminated
Power adapter not properly
connected to auxiliary power
source
Check that power adapter is properly
connected to auxiliary power.
Power adapter not properly
connected to monitor/
defibrillator
Check that power adapter is properly
connected to monitor/defibrillator.
Defective power adapter or
cables
Replace with working power adapter
and cables.
BATTERY CHARGING LED on
monitor/defibrillator not
illuminated
Power adapter not properly
connected to auxiliary power
source or monitor/
defibrillator
Check that power adapter is
connected properly.
Battery not properly inserted
in battery well
Check that battery is properly
inserted in battery well.
Unable to charge battery
with power adapter because
battery charge level is too
low.
Charge battery in Station-Mobile or
REDI-CHARGE battery charger if
available.
Replace battery.
No batteries installed Install at least one battery.
Defective battery Remove battery from service and
replace with working battery.
Unrecognized battery Only use battery that is approved for
use with the LIFEPAK 15 monitor/
defibrillator.
Incompatible power adapter
connected to the monitor/
defibrillator
Only use power adapter that is
approved for use with the
LIFEPAK 15 monitor/defibrillator.
Defective power adapter or
cables
Replace with working power adapter
and cables.
Monitor/defibrillator unable
to recognize installed
battery
Contact qualified service personnel.
CANNOT CHARGE BATTERY
message appears
Defective battery Remove battery from service and
replace with working battery.
Defective power adapter Replace with working power adapter
and cables.
Device unable to charge
battery or batteries
Contact qualified service personnel.
Table 10-2 General Troubleshooting Tips (Continued)
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION