LIFEPAK 15 Monitor/Defibrillator Operating Instructions
General Troubleshooting Tips
General Troubleshooting Tips
If a problem is detected with the LIFEPAK 15 monitor/defibrillator during operation or testing, refer
to the troubleshooting tips in Table 10-2. If the problem cannot be corrected, remove the
LIFEPAK 15 monitor/defibrillator from active use and contact a qualified service technician for
service and repair.
Table 10-2 General Troubleshooting Tips
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION
No power when monitor/
defibrillator is turned
ON
Low battery voltage Replace with fully charged, properly
maintained battery.
Battery connector pin loose,
covered with foreign
substance, or damaged
Remove battery and inspect pins.
Clean if foreign substance present.
Contact a qualified service
technician to replace if bent,
cracked, or loose.
Power adapter not properly
connected to auxiliary power
source
Check that power adapter is properly
connected to auxiliary power.
Power adapter not properly
connected to monitor/
defibrillator
Check that power adapter is properly
connected to monitor/defibrillator.
Defective power adapter or
cables
Replace with working power adapter
and cables.
Defective battery Remove battery from service and
replace with working battery.
ON Led illuminated, but
screen is blank and device
does not operate
Device boot up has failed Press and hold ON until LED turns
off (~5 seconds). Then press
ON to
turn device back on.
If device does not turn off, remove
both batteries and disconnect device
from power adapter, if applicable.
Then reinsert batteries, reconnect
power adapter, and press
ON to turn
device back on.