MAINTAINING THE EQUIPMENT
10
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
10-13
Receiving New Batteries
New batteries do not arrive fully charged. Charge each new battery before use. Batteries may be
charged using any of the following devices:
Station Lithium-ion battery charger for use with the LIFEPAK 15 monitor/defibrillator
Mobile Lithium-ion battery charger for use with the LIFEPAK 15 monitor/defibrillator
REDI-CHARGE battery charger
AC power adapter for use with the LIFEPAK 15 monitor/defibrillator
DC power adapter for use with the LIFEPAK 15 monitor/defibrillator
Storing Batteries
Li-ion batteries self-discharge during storage.
If you store the battery:
Do not remove the Charge Before Use label to indicate that the battery has not yet been
charged.
Store batteries at temperatures between 20° to 25°C (68° to 77°F).
Charge the battery fully within one year of when you receive it. Fully recharge the battery once
per year thereafter.
Charging Batteries
Charge batteries before use. Batteries may be charged in a battery charger, or in the
LIFEPAK 15 monitor/defibrillator if it is connected to an auxiliary power source using a
LIFEPAK 15 monitor/defibrillator power adapter.
CAUTION
POSSIBLE EQUIPMENT DAMAGE
When storing the LIFEPAK 15 monitor/defibrillator for an extended period of time, the
battery should be removed from the device.
WARNING
POSSIBLE LOSS OF POWER DURING PATIENT CARE
Stored batteries lose charge. Failure to charge a stored battery before use may cause
device power failure without warning. Always charge a stored battery before placing it
in active use.