MAINTAINING THE EQUIPMENT
10
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
10-11
7. Confirm that sense markers appear on each QRS complex. If sense markers do not appear, or
appear elsewhere on the ECG, press the
SELECTOR on waveform Channel 1 and adjust ECG
size from the menu.
8. Confirm that the
RATE menu appears.
9. Press
CURRENT and increase the current to 80 mA.
10.Observe the screen for captured complexes. Confirm the
PACER LED flashes with each
delivered pacing pulse.
11.Disconnect the QUIK-COMBO therapy cable from the simulator. Confirm that the pacemaker
stops pacing, the
CONNECT ELECTRODES message appears, and an audible alarm sounds.
12.Reconnect the QUIK-COMBO therapy cable to the simulator. Confirm that the audible alarm
stops, the
PACING STOPPED message is displayed, and current is 0 mA.
13.Wait approximately 30 seconds and confirm that an audible alarm occurs.
14.Increase current to 80 mA. Confirm that audible alarm stops.
15.Press
CHARGE. Confirm that the PACER LED goes off and that heart rate and available energy
are displayed.