POWER ADAPTER
9
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
9-9
Troubleshooting Tips
Table 9-3 Troubleshooting Tips for Power Adapter
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION
POWER LED on power adapter
does not light
Power cord not plugged into
power adapter or power source
Connect power cord.
Defective power adapter or
power cord
Replace with working power
adapter and power cord.
Blown fuse or tripped circuit
breaker in building
Contact qualified service
personnel.
AUXILIARY POWER LED on
monitor/defibrillator not
illuminated
Power adapter not properly
connected to auxiliary power
source or monitor/defibrillator
Check that power adapter is
connected properly.
Defective power adapter or
cables
Replace with working power
adapter and cables.
BATTERY CHARGING LED on
monitor/defibrillator not
illuminated
Power adapter not properly
connected to auxiliary power
source or monitor/defibrillator
Check that power adapter is
connected properly.
Battery not properly inserted in
battery well
Check that battery is properly
inserted in battery well.
Unable to charge battery with
power adapter because battery
charge level is too low
Charge battery in Station-Mobile
or REDI-CHARGE battery charger
if available.
Replace battery.
No batteries installed Install at least one battery.
Defective battery Remove battery from service and
replace with working battery.
Unrecognized battery Only use battery that is approved
for use with the LIFEPAK 15
monitor/defibrillator.
Incompatible power adapter
connected to the monitor/
defibrillator
Only use power adapter that is
approved for use with the
LIFEPAK 15 monitor/defibrillator.
Defective power adapter or
cables
Replace with working power
adapter and cables.
Monitor/defibrillator unable to
recognize installed battery
Contact qualified service
personnel.