Manual

DATA TRANSMISSION
8
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
8-17
Unable to transmit using
a gateway device that has
a functioning direct
connection or Bluetooth
connection
Transmission sites are not set up
in LIFEPAK 15 monitor
Define transmission sites. Each
site name must exactly match the
name of the target device. See
LIFEPAK 15 Monitor/Defibrillator
Setup Options provided with your
device.
Transmission site names in
LIFENET System do not match
site names in LIFEPAK 15
monitor
Check site names in LIFENET
System.
Cellular communication is not
working between the gateway and
transmission sites
Use alternate method to
communicate patient data.
UNABLE TO TRANSMIT
message appears
The LIFEPAK 15 monitor cannot
connect to the device name
selected
Verify target device is ready to
receive transmissions.
Verify target device setup.
Attempt to retransmit.
The output port on the
LIFEPAK 15 monitor is not
configured for the transmission
method you are using
Make sure the transmission site
OUTPUT PORT is configured for the
type of transmission you are
attempting.
Attempt to retransmit.
Target device unable to connect or
unable to connect within timeout
interval
Verify target device is ready to
receive transmissions.
Verify target device setup.
Attempt to retransmit.
The target device requires you to
“accept” incoming
communications
Check your target device for a
required acknowledgment to
connect.
Enter passcode, when prompted.
Set to “Always allow” if possible.
Attempt to retransmit.
Direct connection was disrupted Verify cable connections.
Attempt to retransmit.
TRANSMISSION FAILED
message appears
Computer application program is
not ready or is not available to
receive transmission
Verify target device is running
necessary software.
Attempt to retransmit.
Table 8-3 Troubleshooting Tips for Data Transmission (Continued)
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION