LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Managing Current Patient Records
Managing Current Patient Records
You can add specific patient information to a current Patient Record. For more information, see
"Entering Patient Data" on page 3-24.
How to Print a Current Report
To print a current report:
Select the
SPEED option on this menu to change the speed of the continuous printout when the
PRINT button is pressed. Note that this SPEED option does not affect reports that are printed from
this menu. Available printing speeds for the
PRINT button are:
12.5 MM/SEC
25 MM/SEC
1. Press OPTIONS. The Options menu appears.
2. Select
PRINT. The Options/Print menu
appears.
Options
Patient...
Pacing...
Date / Time...
Alarm Volume...
Archives...
Print...
User Test...
3. If the REPORT, FORMAT, and MODE settings
are correct, select
PRINT. Otherwise, make
changes as desired.
Select a
REPORT:
CODE SUMMARY
TREND SUMMARY
VITAL SIGNS
12-LEAD
Note: A check next to a 12-lead report
indicates that the report was previously
printed.
Select a
FORMAT (for 12-Lead ECG only):
3-CHANNEL
4-CHANNEL
Select a MODE to change the frequency
response of ECG reports:
•MONITOR
DIAGNOSTIC (12-Lead reports always print
in Diagnostic mode)
Options / Print
Print
Report
Format
Mode
Speed
Code Summary
3-Channel
Monitor
25mm/sec