LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Patient Records and Reports
Waveform Events
In addition to being documented in the Event Log, therapy and other selected events also
capture waveform data that are printed with the long and medium CODE SUMMARY report. The
waveform events and the characteristics of waveform data are described in Table 7-3.
Figure 7-2 Possible Event Log Entries
Monitoring
Check patient
Initial rhythm
Replace battery
12-lead
•NIBP
Alarm events
IP label change
Vital signs
5-wire on/off
SpCO/SpMet Advisory
AED
Connect electrodes
Motion
Analysis
Analysis stopped
Shock advised
No shock advised
CPR Metronome
•On/Off
Age-Airway changed
Defibrillation
Manual mode
Charge removed
Shock X, XXXJ
Shock X, Abnormal
Operator Initiated
•Event
Alarms on/off
•Print
VF/VT alarm on/off
Sync on/off
Snapshot
Internal pacer detection on/off
Pacing
Started
•Set
Changed
Stopped
Paused
Transmission
Transmission complete
Transmission failed
Transmission cancelled
Memory Status
Out of waveform memory (memory low)
Out of event memory (memory full)
Table 7-3 Waveform Events
EVENT NAME WAVEFORM DATA (WHEN CAPTURED)
INITIAL RHYTHM 8 seconds after leads on
CHECK PATIENT 8 seconds prior to alert
SHOCK or NO SHOCK ADVISED 2-3 segments of analyzed ECG. Each segment is 2.7 seconds
ANALYSIS X STOPPED 8 seconds of data prior to cessation of analysis