Manual

PADDLE ACCESSORY OPTIONS
6
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
6-9
Figure 6-8 Anterior-Lateral Paddle Position
Anterior-Posterior Placement Place the STERNUM paddle anteriorly over the left precordium and
the
APEX paddle posteriorly behind the heart in the infrascapular area (see Figure 6-9).
Figure 6-9 Anterior-Posterior Paddle Position
Cleaning Pediatric Paddles
Individually protect paddles before and after cleaning to prevent damage to paddle surfaces.
After each use:
1. Wipe or rinse paddle electrodes, cable connector, paddle handles, and cables with mild soap
and water or disinfectant using a damp sponge, towel, or brush. Do not immerse or soak.
2. Dry thoroughly.
3. Examine paddle surfaces, connector, handles, and cables for damage or signs of wear.
Cables that show signs of wear such as loose cable connections, exposed wires, or cable
connector corrosion should be removed from use immediately.
Paddles that have rough or pitted electrodes should be removed from use immediately.
Sternum
Apex
Sternum
Apex