5-38
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Noninvasive Pacing
For general troubleshooting tips, see Table 10-2 on page 10-18.
Pacing starts spontaneously Patient’s heart rate falls below
set pacing rate
Appropriate pacemaker
function; assess patient.
During standby pacing, ECG
lead disconnects and pacing
begins asynchronously
Reconnect ECG lead.
Set pacing rate (ppm) and
ECG paced rate do not
appear to match
Internal error detected Print ECG and calculate the
pace rate.
Improper sensing
(for example, sensing on
T-waves)
QRS complex too small Select another lead.
T-wave too large Adjust ECG size.
SYNC mode will not activate PACER is on. Pacing and Sync
are separate functions and are
not allowed at the same time.
Discontinue pacing, if
appropriate for the patient,
and press
SYNC.
Defibrillator will not turn off Pacemaker is on Turn off
PACER and then press
and hold
ON for at least 2
seconds.
Table 5-4 Troubleshooting Tips for Noninvasive Pacing (Continued)
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION