Manual

THERAPY
5
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
5-37
No muscle response to
pacing
Patient’s heart rate may be
greater than noninvasive pacer
ppm
No action needed.
The Test Load is connected to
therapy cable
Remove the Test Load and
connect therapy electrodes to
cable.
Patient muscle response is
variable and depends on patient
condition. Muscular response to
pacing is not a reliable indicator
of current delivered.
No action needed.
Capture does not occur with
pacing stimulus
Current (mA) set too low Increase pacing current.
(Administer sedation or
analgesia as needed.)
CONNECT CABLE or PACING
STOPPED
message appears
Therapy cable damaged Replace therapy cable and
perform daily checks per
Operator’s Checklist.
CONNECT ELECTRODES
message appears
Pacing cable or electrode
disconnected
Reconnect and set current.
Electrodes not adhering to skin Prepare skin.
Therapy cable damaged Replace therapy cable and
perform daily checks per
Operator’s Checklist.
Electrodes outdated Replace electrodes and set
current.
PACING IN PROGRESS message
appears
CPR pressed Press PACER to stop pacing, if
appropriate, and then press
CPR.
Pacing stops spontaneously
and
PACER FAULT message
appears
Internal error detected Cycle power and start pacing
again.
Obtain service by a qualified
service technician.
Intrinsic QRS complexes not
sensed when pacing
ECG size too low Increase ECG size or select
another lead.
Intrinsic QRS complexes are
occurring during pacemaker's
refractory period
Adjust PPM.
Table 5-4 Troubleshooting Tips for Noninvasive Pacing (Continued)
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION