Manual

5-36
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Noninvasive Pacing
Troubleshooting Tips
Table 5-4 Troubleshooting Tips for Noninvasive Pacing
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION
Device does not function
when
PACER is pressed
Power off Check if power is
ON.
Low battery Replace battery with fully
charged battery.
PACER LED is on, but
CURRENT (mA) will not
increase
Therapy electrodes off Check for message displayed.
Inspect therapy cable and
electrode connections.
PACER LED on,
CURRENT (mA) >0, but
pace markers absent (not
pacing)
Pacing rate set below patient's
intrinsic rate
Increase
PPM.
Pacer oversensing (ECG artifact,
ECG size too high)
Establish clean ECG;
decrease ECG size.
Select nondemand pacing.
Monitor screen displays
distortion while pacing
ECG electrodes not optimally
placed with respect to pacing
electrodes
Reposition electrodes away
from pacing electrodes.
Select another lead (I, II,
or III).
Pacing stops spontaneously
PACER pressed off Press PACER and increase the
current.
Internal error detected. Service
message indicates an internal
failure.
Check for service indicator.
Cycle power and start pacing
again.
Obtain service by a qualified
service technician.
Therapy electrode off Check for message. Check
pacing cable and electrode
connections.
CHARGE pressed Press PACER and increase
current, if pacing desired.
Otherwise, proceed with
defibrillation.
Radio frequency interference Move radio equipment away
from pacemaker.