LIFEPAK 15 Monitor/Defibrillator Operating Instructions
5-31
ABNORMAL ENERGY DELIVERY
message appears and Shock XJ
Abnormal annotated on printout
Open air discharge with
standard paddles
Press paddles firmly on
patient’s chest when
discharging.
Standard paddles placed face-
to-face when
(shock) button
pressed
Perform test discharges per
Operator’s Checklist.
See "Manual Defibrillation
Warnings" on page 5-23.
Patient impedance is out of
range
Increase energy or repeat
shocks as needed.
Consider replacing
disposable therapy
electrodes with new ones.
Internal fault occurred Repeat shock.
Perform CPR and obtain
another defibrillator, if
necessary.
CONNECT ELECTRODES message
appears
Therapy electrodes are not
connected to the therapy cable
Check for electrode
connection.
Electrodes do not adhere
properly to the patient
Press electrodes firmly on
patient’s skin.
Clean, shave, and dry the
patient’s skin as
recommended.
Apply new electrodes.
Electrodes are dry, damaged, or
out of date
Apply new electrodes.
Therapy cable damaged Replace therapy cable and
perform daily checks per
Operator’s Checklist.
REPLACE BATTERY prompt occurs Both batteries are very low Replace one or both
batteries immediately.
Connect to auxiliary power
using approved power
adapter.
CPR time shown (minutes/
seconds) is different than
expected
Metronome is on None. The metronome
adjusts the CPR time to
ensure CPR cycle ends with
compressions.
Incorrect setup option selected Change CPR time setup
option. See LIFEPAK 15
Monitor/Defibrillator Setup
Options provided with your
device
.
Table 5-3 Troubleshooting Tips for Defibrillation and Synchronized Cardioversion (Continued)
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION