LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Automated External Defibrillation (AED)
DISARMING message appears
(energy charge removed)
(shock) button not pressed
within 60 seconds after charge
complete
Recharge the defibrillator, if
desired.
SPEED DIAL pressed Recharge the defibrillator.
Therapy electrodes or cable
disconnected
Reconnect electrode or cable.
Energy did not escalate After a shock, the next
analysis was
NO SHOCK ADVISED
No action needed.
Defibrillator does not
escalate energy when a
NO
SHOCK ADVISED
decision
follows a shock.
Charge time to 360 joules
exceeds 10 seconds
Battery low Replace battery with fully
charged battery.
Connect to auxiliary power
using approved power
adapter.
Operating temperature is too
low
Move patient and device to
warmer environment, if
necessary.
REPLACE BATTERY prompt occurs Both batteries are very low Replace one or both batteries
immediately.
Connect to auxiliary power
using approved power
adapter.
Voice prompts sound faint or
distorted
Low battery power Replace one or both batteries
immediately.
Connect to auxiliary power
using approved power
adapter.
CPR time shown (minutes/
seconds) is different than
expected
Function of metronome None. The metronome
adjusts the CPR time to
ensure CPR cycle ends with
compressions. (See
page 5-14.)
Incorrect setup option
selected
Change CPR time setup
option. See LIFEPAK 15
Monitor/Defibrillator Setup
Options provided with your
device.
Press
CPR and metronome does
not activate
In AED mode, and not in CPR
interval
Wait until CPR interval
(audible “tocks”) to silence
or activate metronome.
Table 5-1 Troubleshooting Tips for AED Mode (Continued)
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION