LIFEPAK 15 Monitor/Defibrillator Operating Instructions
5-19
Troubleshooting Tips
Table 5-1 Troubleshooting Tips for AED Mode
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION
CONNECT ELECTRODES message
appears
Therapy electrodes are not
connected to the therapy cable
Check for electrode
connection.
Electrodes do not adhere
properly to the patient
Press electrodes firmly on
patient’s skin.
Clean, shave, and dry the
patient’s skin as
recommended.
Replace the electrodes.
Electrodes are dry, damaged,
or out of date
Apply new electrodes.
Therapy cable damaged Replace therapy cable and
perform daily checks per
Operator’s Checklist.
CONNECT CABLE message
appears
Therapy cable is disconnected
during charging
Reconnect cable and press
CHARGE again.
Therapy cable damaged Replace therapy cable and
perform daily checks per
Operator’s Checklist.
MOTION DETECTED and STOP
MOTION
messages appear during
analysis
Patient movement Stop CPR during analysis.
When patient is being
manually ventilated, press
ANALYZE after complete
exhalation.
Patient movement because of
agonal respirations
Allow analysis to proceed to
completion—analysis is
delayed no more than
10 seconds due to motion
detection.
Electrical/radio frequency
interference
Move hand-held
communication devices or
other suspected devices away
from the defibrillator, when
possible.
Vehicle motion Stop vehicle during analysis.
Move patient to stable
location, when possible.