LIFEPAK 15 Monitor/Defibrillator Operating Instructions
5-15
AED/Passcode—A passcode screen appears:
Restricted—A
MANUAL MODE DISABLED message appears, an alert tone sounds, and the
LIFEPAK 15 monitor/defibrillator returns to AED mode.
It is important that all users of the LIFEPAK 15 monitor/defibrillator be thoroughly familiar with the
monitor/defibrillator settings and operation before use.
Special AED Setup Options
The following descriptions of AED prompts (voice and text) explain special setup options.
Initial CPR - CPR First
When the INITIAL CPR option is set to CPR FIRST, you are prompted to START CPR immediately after
the AED is turned on, and before an analysis.
Manual Mode
Enter Manual Mode?
No
Push Speed Dial to confirm
Select YES to enter Manual mode.
Yes
Manual Mode
Enter Manual Mode passcode
Rotate and press the SPEED DIAL to enter the
p a s s c o d e .
The code changes to dots to protect the
passcode, and the defibrillator enters Manual
mode.
You have three opportunities to enter the
correct password. After an incorrect attempt,
the message
INCORRECT--TRY AGAIN appears.
After three incorrect attempts, the message
ACCESS DENIED appears, and the defibrillator
returns to AED mode.