LIFEPAK 15 Monitor/Defibrillator Operating Instructions
4-65
Temperature area of home
screen is blank
Initial temperature not
automatically displayed
until device detects
temperature between 24.8°
and 45.2°C (76.6° and
113.4°F)
Allow up to 3 minutes for probe to
equilibrate.
Check that probe is positioned
properly.
Temperature probe not
detected by device
Check connections between probe,
adapter cable, and device.
Check that the sensor is approved
for use with the LIFEPAK 15
monitor/defibrillator.
Contact qualified service personnel.
Table 4-11 Troubleshooting Tips for Temperature Monitoring (Continued)
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION