Manual

MONITORING
4
©2007-2013 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
4-59
No waveform.
No pressure reading.
Transducer closed to
patient
Check patient. Check
stopcock positions and
monitor setup.
Defibrillator shock just
delivered
None.
Invasive BP lower than cuff BP Transducer level higher
than the heart
Reposition transducer to
correct height.
Loose connection Tighten all connections.
Thrombus formation, air
bubbles, or blood in
catheter, kinking, or
arteriospasm
Use syringe to draw back air or
particles in catheter, and then
flush system.
Improper zero reference Open stopcock to air and
rezero transducer.
Defective transducer Replace transducer.
Invasive BP higher than cuff BP Transducer level lower than
the heart
Reposition transducer to
correct height.
Improper zero reference Rezero.
Catheter whip artifact Change catheter tip position.
Use mean pressure values
(mean pressure is less
affected by extremes and will
therefore reflect a more
accurate reading).
Inability to flush system Pressure bag leaking Keep positive pressure in
flush bag at all times.
Remove dressing to check for
external kinking.
Partially kinked or
obstructed catheter
Replace catheter, if clotted.
Inability to zero system Stopcock not open to air or
defective
Check stopcock position.
Replace any defective
stopcocks.
Defective transducer Replace transducer.
System has been zeroed but
continues to indicate zero
reference required
Steps to zero system
performed in wrong order
Close stopcock to air before
placing cap on port.
Table 4-8 Troubleshooting Tips for IP Monitoring (Continued)
OBSERVATION POSSIBLE CAUSE CORRECTIVE ACTION