User Manual

Changing Setup Options
D-4 LIFEPAK 1000 Defibrillator Operating Instructions
Access Energy Protocols from the AED menu. The underlined bold options in Ta bl e D - 5 are the
factory default settings.
Access CPR Setup from the AED menu. The underlined bold options in Tab l e D- 6 are the factory
default settings.
Ta b l e D- 5 AED Setup Menu—Energy Protocols Submenu
Menu Item Help Message Options
ENERGY 1
Select energy level for shock 1. 150, 175, 200, 225, 250, 275, 300, 325,
360
*
joules.
*
When selecting 360 joules for energy 1, consider AED use in children.
ENERGY 2
Select energy equal to or
greater than Energy 1.
150, 175, 200, 225, 250, 275, 300, 325,
360 J
oules.
ENERGY 3
Select energy equal to or
greater than Energy 2.
150, 175, 200, 225, 250, 275, 300, 325,
360 joules.
FLEXIBLE
PROTOCOL
Repeat previous energy after
NO SHOCK ADVISED (only
when
NO SHOCK ADVISED
follows a shock).
ON, OFF.
STACKED
SHOCKS
Enable consecutive shocks
without CPR.
ON, OFF.
PREVIOUS PAGE
Go back to previous page.
Ta b l e D- 6 AED Setup Menu—CPR Submenu
Menu Item Help Message Options
CPR TIME 1
Set CPR interval after shocks. 15, 30, 45, 60, 90, 120, 180 seconds.
CPR TIME 2
Set CPR interval after NO
SHOCK ADVISED
.
15, 30, 45, 60, 90, 120, 180 seconds.
INITIAL CPR
Enable Initial CPR. OFF, ANALYZE FIRST, CPR FIRST.
INITIAL CPR TIME
Set CPR interval after first
analysis.
15, 30, 45, 60, 90, 120, 180 seconds.
PRESHOCK CPR
Set CPR interval before
SHOCK ADVISED decisions.
OFF, 15, 30 seconds.
CPR PROMPT
Enable extended CPR prompt:
PROVIDE RESCUE BREATHS
AND CHEST COMPRESSIONS
ON
, OFF.
PREVIOUS PAGE