cprMAX™ Technology
LIFEPAK 1000 Defibrillator Operating Instructions C-3
©2006-2012 Physio-Control, Inc.
Pulse Check
The PULSE CHECK option inserts prompting to check for a pulse or check the patient,
depending on the
PULSE PROMPT setting. The choices for PULSE CHECK are: ALWAYS,
AFTER EVERY NSA, AFTER SECOND NSA
, and NEVER. The default setting is NEVER.
•The
ALWAYS option prompts for a pulse check after CPR TIME 1 and 2, after a NO SHOCK
ADVISED
decision, after a single SHOCK ADVISED decision with STACKED SHOCKS OFF, or
after three consecutive
SHOCK ADVISED decisions if STACKED SHOCKS is ON.
•The
AFTER EVERY NSA option prompts for a pulse check after every NO SHOCK ADVISED
decision.
•The
AFTER SECOND NSA option prompts for a pulse check after the second analysis if the
second analysis results in a
NO SHOCK ADVISED decision, regardless of the first analysis
decision (
SHOCK ADVISED or NO SHOCK ADVISED).
•The
NEVER option eliminates all PULSE CHECK prompts.