14 TrueCPR Device Instructions for Use
Cleaning the Device
Cleaning the Device
The TrueCPR device should be cleaned and inspected after each use.
1 Clean the device with a sponge or cloth that has been dampened with one of the cleaning agents
shown below:
Quaternary ammonium compounds
Hydrogen peroxide, 3% solution
Sodium dichloroisocyanurate (NaDCC), 3000 ppm solution
Chlorine bleach, 1:10 dilution with water
2 Inspect the device and cable for wear or damage. Remove from service if damaged.
3 After cleaning, wipe off solution with a clean, damp cloth.
IMPORTANT! Do not immerse or soak the chest pad or back pad.
Storing the Device
Place chest pad onto back pad. Store in carrying bag so that battery readiness indicator is visible.
CAUTION
POSSIBLE DEVICE DAMAGE: If the TrueCPR device will be stored for an extended period of time,
remove the batteries to prevent possible damage due to battery leakage.