Item Brochure
©2012 Physio-Control, Inc. 13
Data Management
Data Management
The TrueCPR device can store compression data for three 60-minute sessions, or up to six sessions
totaling 180 minutes. When all available memory has been used, the data from the oldest use is
overwritten automatically.
Data can be transferred to a computer with TrueCPR device-compatible software via USB connection.
Event reports can be printed directly from the software. For information on TrueCPR device reports and
full event review functionality, contact your local Physio-Control representative.
To initiate data transfer:
1 Connect a USB cable between the TrueCPR device and a computer with compatible software.
Note: Use only a USB cable that meets the following specifications: USB 2.0 A-Male to Mini-B 5-pin
Male, 28/24 AWG cable with ferrite core (gold plated), 1 m (3 ft).
2 Turn on the TrueCPR device. If the software on the computer does not start automatically, open the
program manually.
3 Follow the instructions in the software on the computer to select the TrueCPR device and initiate data
transfer. Data transfer may take up to 5 minutes if memory is full. The following screens may appear on
the TrueCPR device:
Troubleshooting Tips for Data Management
WARNINGS
SHOCK HAZARD: Do not use the USB connector if the TrueCPR device is damaged.
POSSIBLE ELECTRICAL INTERFERENCE: Using cables or accessories that are not specified for use
with the TrueCPR device may result in increased emissions or decreased immunity from
electromagnetic or radio frequency interference (RFI) for this device. This could affect the performance
of the TrueCPR device or equipment in close proximity. Use only cables and accessories specified in
these instructions for use.
Data transfer in progress. Data transfer successful. Data transfer failed.
OBSERVATION CORRECTIVE ACTION
Feedback screen appears on
TrueCPR device and metronome is
heard.
• Confirm USB cable is properly connected.
• Confirm computer is on.
• Replace USB cable.
• Use recommended USB cable (see Step 1 on page 13).
Data Transfer Failed
screen appears. • Turn TrueCPR device off and check connections. Then turn
device on and attempt data transfer again.
• If problem persists, contact Technical Support.