Item Brochure

©2012 Physio-Control, Inc. 11
Troubleshooting Tips
Troubleshooting Tips
These troubleshooting tips provide immediate corrective actions for possible problems. If a problem
persists, contact your local Physio-Control representative for assistance. In the USA, call Technical
Support at 800.442.1142.
OBSERVATION CORRECTIVE ACTION
No audible metronome. Use compression rate meter for guidance.
Device does not function. Only use device within specified operating temperature range
of 0°–40°C (32°–104°F).
Remove batteries, wait 30 seconds, and then re-install
batteries.
Perform CPR unaided.
Display is visible, but does not provide
compression feedback.
Turn device off and back on. Do not apply pressure to chest
pad until calibration is complete and feedback screen
appears.
Perform CPR unaided.
Alert indicator persists. Turn device off and back on. If indicator persists, perform
CPR unaided.
Contact Technical Support or your local Physio-Control
representative for assistance.