29
Most probable cause
Battery exhausted.
Extraneous cause:
radio antenna or similar
interfering devices nearby.
a) Hearing instrument set
to wrong program.
b) Receiver switched OFF.
c) Radio frequency not
synchronized.
EasyLink+ is positioned too
far away from the speaker.
Remedy
Charge the EasyLink+
(2 hours minimum).
Change channel
(see page 25).
a) Check hearing
instrument program.
b) Switch ON receiver.
c) Switch EasyLink+ OFF
and ON by holding it
close to the FM receivers.
Move the EasyLink+
closer to the speaker.