Datasheet
https://www.phoenixcontact.com/us/produkte/2865586
Nov 14, 20132/8
Temperature measuring transducer-MACX MCR-EX-SL-TC-I-
NC-2865586
Technical data
Ambient conditions
Maximum altitude
≤ 2000 m
Permissible humidity (operation)
5 % ... 95 % (no condensation)
Degree of protection
IP20
Input data
Input
Intrinsically safe
Sensor types that can be used (TC)
E, J, K, N as per IEC / EN 60584, L as per DIN 43760
Temperature measuring range
-250 °C ... 1372 °C (Range depending on the sensor type)
Input signal range
-20 mV ... 70 mV
Measuring range span
(Min. 50 K for thermocouples, 3 mV for mV sources)
Output data
Signal output
Current output
Current output signal
0 mA ... 20 mA
Current output signal
4 mA ... 20 mA
Load/output load current output
max. 500 Ω
Output ripple (current)
< 50 µA
PP
Behavior in the event of a sensor error
As per NE 43 or can be freely defined
Power supply
Nominal supply voltage
24 V DC
Supply voltage range
19.2 V DC ... 30 V DC
Max. current consumption
< 40 mA (24 V DC)
Power consumption
< 1 W
Connection data
Conductor cross section solid min.
0.2 mm²
Conductor cross section solid max.
2.5 mm²
Conductor cross section stranded min.
0.2 mm²
Conductor cross section stranded max.
2.5 mm²
Conductor cross section AWG/kcmil min.
24
Conductor cross section AWG/kcmil max
14
Stripping length
7 mm
Screw thread
M3
Connection method
Screw connection
Tightening torque, min
0.5 Nm
Tightening torque max
0.6 Nm
General
No. of channels
1
Temperature coefficient, typical
0.01 %/K
Typical cold point errors
± 1 K
Step response (0–99%)
typ. 800 ms (With SIL)