Datasheet

https://www.phoenixcontact.com/us/produkte/2864286
Nov 13, 20133/9
Resistance thermometer measuring transducer-MINI MCR-SL-
PT100-UI-SP-NC-2864286
Technical data
Connection data
Conductor cross section stranded max.
2.5 mm²
Conductor cross section AWG/kcmil min.
24
Conductor cross section AWG/kcmil max
12
Stripping length
8 mm
General
No. of channels
1
Maximum temperature coefficient
< 0.02 %/K
Protective circuit
Transient protection
Electrical isolation
Basic insulation according to EN 61010
Surge voltage category
II
Pollution degree
2
Rated insulation voltage
50 V AC/DC
Test voltage, input/output/supply
1.5 kV (50 Hz, 1 min.)
Electromagnetic compatibility
Conformance with EMC Directive 2004/108/EC
Noise emission
EN 61000-6-4
Noise immunity
EN 61000-6-2 When being exposed to interference, there may be
minimal deviations.
Color
green
Housing material
PBT
Mounting position
Any
Assembly instructions
The T connector can be used to bridge the supply voltage. It can
be snapped onto a 35 mm DIN rail according to EN 60715.
Conformance
CE-compliant
ATEX
 II 3 G Ex nA IIC T4 Gc X
UL, USA / Canada
UL 508 Recognized
UL, USA / Canada
Class I, Div. 2, Groups A, B, C, D T5
GL
GL EMC 2 D
EMC data
Name
Electromagnetic RF field
Standards/regulations
EN 61000-4-3
Typical deviation from the measuring range final value
10 %
Name
Fast transients (burst)
Standards/regulations
EN 61000-4-4
Typical deviation from the measuring range final value
10 %
Name
Conducted interferences
Standards/regulations
EN 61000-4-6
Typical deviation from the measuring range final value
10 %