Service manual

Section 6: Disassembly Guide
6-7
Procedure
d) Disconnect large ribbon-cable connector from main PCB.
e) Unscrew NIBP tubing connector from pump to main PCB. Front and rear
case assemblies are now completely separable from one another.
Illustration
Step B2
Procedure
To remove SpO2 module, ECG/Temp connector assembly, and encoder
assembly:
SpO
2 module: remove three screws located at corners of SpO2 module,
lift off ground wire and foil shield, pull SpO
2 module straight up to
disengage it from main PCB.
ECG/Temp cable/connector assembly: unplug connector from Main
PCB, remove screws fastening ECG connector to front case, rotate
temperature connector counter-clockwise to unscrew it from fastening nut
on outside of front case.
The encoder assembly: unplug connector from main PCB. Remove
knob, as described in step A1. Use 9/16hex socket to unscrew fastening
nut on outside of front case. Encoder may be pulled away from front case.
SpO2 module